The most used temperature for calibration is 0°C.
The normal way of creating 0°C is via a mixture of ice and water in a Dewar Flask.
However, this can give errors of up to 4°C because water is densest at 4°C and so as the ice melts the temperatures at the bottom of the flask can rise to 4°C.
In the design of the ice flask offered by Isothermal Technology Ltd., these problems have been eliminated by stirring the water/ice mixture and segregating the ice from the water in the measuring zone.
This stirred ice/water bath is designed and built according to National Laboratory recommendations.
Using demineralised water, accuracies of ±0.005K are obtainable. Typically the bath will last for 4 hours before recharging with ice.
The ice is contained around and below the compartment where up to 4 probes can be placed for calibration or referencing purposes.
An option permits a water triple point cell to be maintained within the stirred ice bath.
Pirkti Isotech 813 - simple stirred ice/water bath
Jei turite klausimų dėl įrangos, susisiekite su mumis elektroniniu paštu: This email address is being protected from spambots. You need JavaScript enabled to view it. ARBA telefonu: +370 (5) 210 22 74.
Isotech 813 - simple stirred ice/water bath, kaina: ---- eurų